Use of a disposable water filter for prevention of false-positive results due to nontuberculosis mycobacteria in a clinical laboratory performing routine acid-fast staining for tuberculosis

Hui Z. Tu, Chiao Shan Chen, Tsi S. Huang, Wen Kuei Huang, Yao Shen Chen, Yung Ching Liu, Yusen Eason Lin

研究成果: 雜誌貢獻文章同行評審

6 引文 斯高帕斯(Scopus)

摘要

A point-of-use 0.2-μm filter was evaluated for elimination of nontuberculosis mycobacteria in laboratory water to reduce false-positive acid-fast bacillus staining results. Use of the point-of-use filter can significantly reduce the false-positive rate to 1.2% compared to samples treated with tap water (10.7%) and deionized water (8.7%).