TY - JOUR
T1 - Trueness of removable partial denture frameworks additively manufactured with selective laser melting
AU - Peng, Pei Wen
AU - Hsu, Ching Ying
AU - Huang, Huei Yu
AU - Chao, Jen Chih
AU - Lee, Wei Fang
N1 - Funding Information:
Supported by the research grant from Taipei Medical University , grant# TMU105-AE1-B40 , and in part by the Ministry of Science and Technology , grant# 106-2314-B-038-019 .
Publisher Copyright:
© 2020 Editorial Council for the Journal of Prosthetic Dentistry
Copyright:
Copyright 2020 Elsevier B.V., All rights reserved.
PY - 2022/1
Y1 - 2022/1
N2 - Statement of problem: Although studies have reported on selective laser melting (SLM)-fabricated removable partial dentures (RPDs), research addressing the trueness of SLM-fabricated RPD metal frameworks is sparse. Purpose: The purpose of this in vitro study was to evaluate the trueness of powdered cobalt-chromium (Co-Cr) or titanium-6 aluminum-4 vanadium (Ti-6Al-4V) alloy frameworks for RPDs fabricated by SLM. Material and methods: A digital scan of a Kennedy class II mandible typodont was obtained to design an RPD framework by using a computer-aided design (CAD) software program (denoted as CRF). Two experimental groups of frameworks were fabricated from the CRF by using SLM in alloys of Co-Cr (SLM-Co-Cr, n=6) and in Ti-6Al-4V (SLM-Ti-6Al-4V, n=6) while a control group was fabricated by using traditional lost-wax casting following stereolithography (CA-Co-Cr, n=6). In total, 18 RPD frameworks were digitally scanned (denoted as CRF), with each scan then superimposed on the CRF and evaluated for discrepancies by using a 3D analysis software program. A nonparametric Kruskal-Wallis test was performed to determine differences in trueness among groups (α=.05). Results: The CA-Co-Cr group showed the highest discrepancy between CEF and CRF. Statistically significant differences were found between the CA-Co-Cr and SLM-fabricated groups (P=.03 for Co-Cr, and P=.016 for Ti-6Al-4V). However, no significant difference was found between the SLM-Co-Cr and SLM-Ti-6Al-4V groups (P=.787). Conclusions: SLM-fabricated RPD frameworks exhibited higher trueness than CA-Co-Cr fabricated ones, indicating the potential of selective laser melting to produce the geometric shapes required for accurate dental restorations.
AB - Statement of problem: Although studies have reported on selective laser melting (SLM)-fabricated removable partial dentures (RPDs), research addressing the trueness of SLM-fabricated RPD metal frameworks is sparse. Purpose: The purpose of this in vitro study was to evaluate the trueness of powdered cobalt-chromium (Co-Cr) or titanium-6 aluminum-4 vanadium (Ti-6Al-4V) alloy frameworks for RPDs fabricated by SLM. Material and methods: A digital scan of a Kennedy class II mandible typodont was obtained to design an RPD framework by using a computer-aided design (CAD) software program (denoted as CRF). Two experimental groups of frameworks were fabricated from the CRF by using SLM in alloys of Co-Cr (SLM-Co-Cr, n=6) and in Ti-6Al-4V (SLM-Ti-6Al-4V, n=6) while a control group was fabricated by using traditional lost-wax casting following stereolithography (CA-Co-Cr, n=6). In total, 18 RPD frameworks were digitally scanned (denoted as CRF), with each scan then superimposed on the CRF and evaluated for discrepancies by using a 3D analysis software program. A nonparametric Kruskal-Wallis test was performed to determine differences in trueness among groups (α=.05). Results: The CA-Co-Cr group showed the highest discrepancy between CEF and CRF. Statistically significant differences were found between the CA-Co-Cr and SLM-fabricated groups (P=.03 for Co-Cr, and P=.016 for Ti-6Al-4V). However, no significant difference was found between the SLM-Co-Cr and SLM-Ti-6Al-4V groups (P=.787). Conclusions: SLM-fabricated RPD frameworks exhibited higher trueness than CA-Co-Cr fabricated ones, indicating the potential of selective laser melting to produce the geometric shapes required for accurate dental restorations.
UR - http://www.scopus.com/inward/record.url?scp=85096598229&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85096598229&partnerID=8YFLogxK
U2 - 10.1016/j.prosdent.2020.06.035
DO - 10.1016/j.prosdent.2020.06.035
M3 - Article
AN - SCOPUS:85096598229
SN - 0022-3913
VL - 127
SP - 122
EP - 127
JO - Journal of Prosthetic Dentistry
JF - Journal of Prosthetic Dentistry
IS - 1
ER -