Thermal analysis on monopolar electrosurgery device with novel electrode for minimizing tissue damage using computer-aided system

Yung Sheng Yen, Hung Ta Lin, Han Yi Cheng

研究成果: 雜誌貢獻文章同行評審

摘要

The aim of the present study was to investigate the thermal injury in the liver after a minimally invasive electrosurgery with a titanium dioxide (TiO2), a chromium nitride (CrN) and a diamondlike carbon (DLC) surface modifications. Three-dimensional (3-D) liver models were reconstructed using magnetic resonance imaging to simulate the electrosurgical procedure. Several reports have investigated 3D finite element models for electrosurgical devices, but few have examined a model for devices with treated layers. Data indicated that the temperature decreased significantly when minimally electrosurgery with nanostructured TiO2, CrN and DLC thin films were used. Moreover, DLC electrosurgery created a relatively small thermal injury area and decreased 12.93% temperature compared with control groups. The present study reveals that the novel nanostructured thin film on electrode substrates is an effective means of improving the performance of reducing excessive thermal injury and uniformly distributing temperature in the liver.

原文英語
頁(從 - 到)883-889
頁數7
期刊Journal of Biomaterials and Tissue Engineering
6
發行號11
DOIs
出版狀態已發佈 - 11月 2016