Stress analysis of different wall thicknesses of implant fixture with various boundary levels

Haw-Ming Huang, Li De Wu, Sheng-Yang Lee, H. C. Chen, Junn Lain Chen, Chien-Tsu Chen

研究成果: 雜誌貢獻文章同行評審

3 引文 斯高帕斯(Scopus)

摘要

The aim of the present work is to develop 3D finite element models of implant fixture with different wall thicknesses to predict maximum stress concentration sites and distribution contours after loading. A maximum lateral force of 150 N was applied to simulate horizontal occlusal forces. When the fixtures were constrained to simulate different boundary levels, the maximum equivalent stress (max EQV) was always located at the implant-bone interface. Max EQV increased when the wall thickness or boundary level was reduced to a certain extent. The fixture with a wall thickness of 0.97 mm demonstrated the smallest stress increase ratio when the boundary level was lowered. Our results indicated that both wall thickness and the boundary level played important roles in maintaining a well-distributed stress level within the fixture. The stress concentration decreased when the fixture wall became thicker, however, this effect was less significant when the surrounding bone level was reduced.
原文英語
頁(從 - 到)267-272
頁數6
期刊Journal of Medical Engineering and Technology
24
發行號6
DOIs
出版狀態已發佈 - 2000

ASJC Scopus subject areas

  • 生物醫學工程

指紋

深入研究「Stress analysis of different wall thicknesses of implant fixture with various boundary levels」主題。共同形成了獨特的指紋。

引用此