5 引文 斯高帕斯(Scopus)

摘要

Glutamine, glutamate and glutathione are key modulators of excessive oxidative stress in tumor cells. In this study, we developed a rapid and accurate HILIC-MS/MS method to simultaneously determine concentrations of cellular glutamine, glutamate and glutathione. A bared silica HILIC column was employed to analyze these polar metabolites. The LC-MS parameters were optimized to achieve high sensitivity and selectivity. The analysis can be completed within 4 min under optimal conditions. The method was validated in terms of accuracy, precision, and linearity. Intra-day (n = 9) precision was within 2.68–6.24% among QCs. Inter-day precision (n = 3) was below 12.4%. The method accuracy was evaluated by the recovery test, and the accuracy for three analytes were between 91.6 and 110%. The developed method was applied to study antioxidant function of GLS2 in non-small cell lung cancer cells. Changes in concentrations of glutamine, glutamate and glutathione revealed that the overexpression of GLS2 could effectively decrease oxidative stress. In summary, this study developed a rapid HILIC-MS/MS method for quantification of GLS2-related metabolites that could facilitate elucidation of the role of GLS2 in tumor development.

原文英語
頁(從 - 到)39-44
頁數6
期刊Analytical Biochemistry
539
DOIs
出版狀態已發佈 - 12月 15 2017

ASJC Scopus subject areas

  • 生物物理學
  • 生物化學
  • 分子生物學
  • 細胞生物學

指紋

深入研究「Rapid quantification of glutaminase 2 (GLS2)-related metabolites by HILIC-MS/MS」主題。共同形成了獨特的指紋。

引用此