Quantitative evaluation of thrombosis by electrochemical methodology

Chun Che Shih, Chun Ming Shih, Yea Yang Su, Lin Hui Julie Su, Mau Song Chang, Shing Jong Lin

研究成果: 雜誌貢獻文章同行評審

22 引文 斯高帕斯(Scopus)

摘要

Electrochemical reactions between blood and metal electrodes have been studied since 1928. Little is known about the actual current density induced during the reaction. In this study, an in situ continuous monitoring method was developed to detect the progress of thrombosis on an oxidized 316L stainless steel coil was deployed inside the artery. Three stages of current density were observed on a 316L wire passivated with polycrystalline oxide film. In contrast, no significant current density was detected for a wire passivated with amorphous oxide film. Results showed that this in situ electrochemical monitoring method is sensitive to the passivated film on the stainless steel coil and could provide efficient and reliable information on the control of thrombosis on cardiovascular devices.
原文英語
頁(從 - 到)103-109
頁數7
期刊Thrombosis Research
111
發行號1-2
DOIs
出版狀態已發佈 - 2003

ASJC Scopus subject areas

  • 血液學

指紋

深入研究「Quantitative evaluation of thrombosis by electrochemical methodology」主題。共同形成了獨特的指紋。

引用此