Preparation of radioiodinated secretin for radioimmtmoassay

Chung-Huei Hsu, Kuo Wei Yin, Tsai Hsia Hong, Tswen Kei Tang, Long Shyong Lee, Tong H. Chang, Jean-Dean Liu, P. H. Chen

研究成果: 雜誌貢獻文章同行評審

摘要

Radioiodination of synthetic human secretin on its N-terminal histidyl residue was not difficult when a greater amount of Chloramine T and a longer reaction time were employed to achieve better incorporation of125I. The radioiodinated tracer for an optimal radioimmunoassay required purification. The combination of Sep-pak C18 Cartridge and high performance liquid chromatography for the purification of125I-secretin in our study revealed that the Sep-pak cartridge was a preliminary step in removing unlabeled radioactive iodide, the reactant, and labeled materials unadsorbed to the cartridge. The eluate eluted from the Sep-pak containing high radioactivity and high immunoreactivity to the antibody were selected for further purification by HPLC which eliminated undesirable radiolabeled substances with lower immunoreactivity. The purified radiolabeled secretin was used in developing a sensitive radioimmunoassay.
原文英語
頁(從 - 到)83-87
頁數5
期刊Annals of Nuclear Medicine
5
發行號3
DOIs
出版狀態已發佈 - 9月 1991
對外發佈

ASJC Scopus subject areas

  • 放射學、核子醫學和影像學

指紋

深入研究「Preparation of radioiodinated secretin for radioimmtmoassay」主題。共同形成了獨特的指紋。

引用此