TY - JOUR
T1 - Micro/nanostructured surface modification using femtosecond laser pulses on minimally invasive electrosurgical devices
AU - Lin, Chia Cheng
AU - Lin, Hao Jan
AU - Lin, Yun Ho
AU - Sugiatno, Erwan
AU - Ruslin, Muhammad
AU - Su, Chen Yao
AU - Ou, Keng Liang
AU - Cheng, Han Yi
N1 - Publisher Copyright:
© 2016 Wiley Periodicals, Inc.
PY - 2017/5/1
Y1 - 2017/5/1
N2 - The purpose of the present study was to examine thermal damage and a sticking problem in the tissue after the use of a minimally invasive electrosurgical device with a nanostructured surface treatment that uses a femtosecond laser pulse (FLP) technique. To safely use an electrosurgical device in clinical surgery, it is important to decrease thermal damage to surrounding tissues. The surface characteristics and morphology of the FLP layer were evaluated using optical microscopy, scanning electron microscopy, and transmission electron microscopy; element analysis was performed using energy-dispersive X-ray spectroscopy, grazing incidence X-ray diffraction, and X-ray photoelectron spectroscopy. In the animal model, monopolar electrosurgical devices were used to create lesions in the legs of 30 adult rats. Animals were sacrificed for investigations at 0, 3, 7, 14, and 28 days postoperatively. Results indicated that the thermal damage and sticking situations were reduced significantly when a minimally invasive electrosurgical instrument with an FLP layer was used. Temperatures decreased while film thickness increased. Thermographic data revealed that surgical temperatures in an animal model were significantly lower in the FLP electrosurgical device compared with that in the untreated one. Furthermore, the FLP device created a relatively small area of thermal damage. As already mentioned, the biomedical nanostructured layer reduced thermal damage and promoted the antisticking property with the use of a minimally invasive electrosurgical device.
AB - The purpose of the present study was to examine thermal damage and a sticking problem in the tissue after the use of a minimally invasive electrosurgical device with a nanostructured surface treatment that uses a femtosecond laser pulse (FLP) technique. To safely use an electrosurgical device in clinical surgery, it is important to decrease thermal damage to surrounding tissues. The surface characteristics and morphology of the FLP layer were evaluated using optical microscopy, scanning electron microscopy, and transmission electron microscopy; element analysis was performed using energy-dispersive X-ray spectroscopy, grazing incidence X-ray diffraction, and X-ray photoelectron spectroscopy. In the animal model, monopolar electrosurgical devices were used to create lesions in the legs of 30 adult rats. Animals were sacrificed for investigations at 0, 3, 7, 14, and 28 days postoperatively. Results indicated that the thermal damage and sticking situations were reduced significantly when a minimally invasive electrosurgical instrument with an FLP layer was used. Temperatures decreased while film thickness increased. Thermographic data revealed that surgical temperatures in an animal model were significantly lower in the FLP electrosurgical device compared with that in the untreated one. Furthermore, the FLP device created a relatively small area of thermal damage. As already mentioned, the biomedical nanostructured layer reduced thermal damage and promoted the antisticking property with the use of a minimally invasive electrosurgical device.
KW - electrosurgical instrument
KW - laser surface treatment
KW - nanostructured film
KW - thermal damage
UR - http://www.scopus.com/inward/record.url?scp=84956606124&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84956606124&partnerID=8YFLogxK
U2 - 10.1002/jbm.b.33613
DO - 10.1002/jbm.b.33613
M3 - Article
C2 - 26821850
AN - SCOPUS:84956606124
SN - 1552-4973
VL - 105
SP - 865
EP - 873
JO - Journal of Biomedical Materials Research - Part B Applied Biomaterials
JF - Journal of Biomedical Materials Research - Part B Applied Biomaterials
IS - 4
ER -