Label-Free Cholesteric Liquid Crystal Biosensing Chips for Heme Oxygenase-1 Detection within Cerebrospinal Fluid as an Effective Outcome Indicator for Spontaneous Subarachnoid Hemorrhage

Hui Tzung Luh, Yi Wei Chung, Po Yi Cho, Yu Cheng Hsiao

研究成果: 雜誌貢獻文章同行評審

3 引文 斯高帕斯(Scopus)

摘要

A novel device for cholesteric liquid crystal (LC; CLC)-based biosensing chips for detecting heme oxygenase (HO)-1 within the cerebrospinal fluid (CSF) was invented. In the CLC device, the reorientation of the LCs was strongly influenced by the alignment layer surface and adjacent LCs. When the substrate was coated with the alignment layer, the CLCs oriented homeotropically in a focal conic state. Once HO-1 was immobilized onto the orientation sheet-coated substrate, the CLC changed from a focal conic state to a bright planar state by disrupting the CLCs. The concentration of HO-1 within CSF was shown to be an effective outcome indicator for patients with a spontaneous subarachnoid hemorrhage. We showed that the CLC immunoassaying can be used to measure HO-1 with a lower detection limit of about 10 ng/mL. The linear range was 10 ng/mL to 1 mg/mL. An easy-to-use, rapid-detection, and label-free CLC immunoassay device is proposed.

原文英語
文章編號204
期刊Biosensors
12
發行號4
DOIs
出版狀態已發佈 - 4月 2022