Evaluation of Clinical Application and Dosimetric Comparison of Treatment Plans of Gamma Knife and CyberKnife in Treating Arteriovenous Malformations

Chun Yuan Kuo, Yi Chieh Tsai, An Cheng Shiau, Hao Wen Cheng, Hsiao Wei Yu, Yu Kai Su, Jo Ting Tsai

研究成果: 雜誌貢獻文章同行評審

3 引文 斯高帕斯(Scopus)

摘要

Purpose: To analyze and compare the characteristics of dose distributions for Leksell Gamma Knife Perfexion (LGK-PFX) and CyberKnife (CK) in treating arteriovenous malformations (AVMs). Subjects and Methods: Twenty-four patients with AVMs who received CK radiosurgery at a prescribed dose (PD) of 16-25 Gy in a single fraction were selected. A LGK-PFX treatment plan with the same PD was designed for each patient. Dosimetric values for both systems were compared with respect to the conformity index (CI); selectivity index (SI); gradient index (GI) of 75, 50, and 25% of the PD; heterogeneity index; volume of the brain tissue covered by doses of 10 and 12 Gy; maximum dose delivered to the brainstem; and beam-on time. Results: The CIs of LGK-PFX and CK were 0.744 ± 0.075 and 0.759 ± 0.071 (p = 0.385), respectively. The SIs of LGK-PFX and CK were 0.764 ± 0.081 and 0.780 ± 0.076 (p = 0.424), respectively. The GI75%, GI50%, and GI25% values of LGK-PFX and CK were 1.028 ± 0.123 and 2.439 ± 0.338 (p < 0.001), 3.169 ± 0.265 and 4.972 ± 0.852 (p < 0.001), and 8.650 ± 0.914 and 14.261 ± 2.476 (p < 0.001), respectively. Volumes of the brain tissue covered by 10 Gy and 12 Gy for LGK-PFX and CK (p < 0.001) exhibited a significant difference. Conclusions: LGK-PFX and CK exhibited similar dose conformity. LGK-PFX showed superior normal tissue sparing.
原文英語
頁(從 - 到)142-148
頁數7
期刊Stereotactic and Functional Neurosurgery
95
發行號3
DOIs
出版狀態已發佈 - 7月 1 2017

ASJC Scopus subject areas

  • 手術
  • 神經病學(臨床)

指紋

深入研究「Evaluation of Clinical Application and Dosimetric Comparison of Treatment Plans of Gamma Knife and CyberKnife in Treating Arteriovenous Malformations」主題。共同形成了獨特的指紋。

引用此