摘要
Purpose: To analyze and compare the characteristics of dose distributions for Leksell Gamma Knife Perfexion (LGK-PFX) and CyberKnife (CK) in treating arteriovenous malformations (AVMs). Subjects and Methods: Twenty-four patients with AVMs who received CK radiosurgery at a prescribed dose (PD) of 16-25 Gy in a single fraction were selected. A LGK-PFX treatment plan with the same PD was designed for each patient. Dosimetric values for both systems were compared with respect to the conformity index (CI); selectivity index (SI); gradient index (GI) of 75, 50, and 25% of the PD; heterogeneity index; volume of the brain tissue covered by doses of 10 and 12 Gy; maximum dose delivered to the brainstem; and beam-on time. Results: The CIs of LGK-PFX and CK were 0.744 ± 0.075 and 0.759 ± 0.071 (p = 0.385), respectively. The SIs of LGK-PFX and CK were 0.764 ± 0.081 and 0.780 ± 0.076 (p = 0.424), respectively. The GI75%, GI50%, and GI25% values of LGK-PFX and CK were 1.028 ± 0.123 and 2.439 ± 0.338 (p < 0.001), 3.169 ± 0.265 and 4.972 ± 0.852 (p < 0.001), and 8.650 ± 0.914 and 14.261 ± 2.476 (p < 0.001), respectively. Volumes of the brain tissue covered by 10 Gy and 12 Gy for LGK-PFX and CK (p < 0.001) exhibited a significant difference. Conclusions: LGK-PFX and CK exhibited similar dose conformity. LGK-PFX showed superior normal tissue sparing.
原文 | 英語 |
---|---|
頁(從 - 到) | 142-148 |
頁數 | 7 |
期刊 | Stereotactic and Functional Neurosurgery |
卷 | 95 |
發行號 | 3 |
DOIs | |
出版狀態 | 已發佈 - 7月 1 2017 |
ASJC Scopus subject areas
- 手術
- 神經病學(臨床)