摘要
Atrial rate variations (ARV) of nine atrioventricular (AV) block patients loaded with VDD pacemakers were analyzed with pacemaker-off mode, programmed at different fixed ventricular pacing rates (60, 90, 120 bpm) and with different AV delays in VDD pacing (80, 120, 160, 200 ms) to study the influences of ventricular rate and AV delay. ARV were measured by PP intervals on the atrial electrocardiogram (ECG) that was non-invasively recorded from an esophageal lead. In results, all ARV measures had no significant difference under different AV delays. Although no significant difference existed among all ARV frequency domain measures, the locations of the major powers were very diverse for different individuals under different pacing rates. However, during pacemaker-off mode, the major spectral powers of ARV were centered on the band above 0.4 Hz.
原文 | 英語 |
---|---|
主出版物標題 | Computers in Cardiology |
發行者 | IEEE |
頁面 | 555-557 |
頁數 | 3 |
出版狀態 | 已發佈 - 1999 |
對外發佈 | 是 |
事件 | The 26th Annual Meeting: Computers in Cardiology 1999 - Hannover, Ger 持續時間: 9月 26 1999 → 9月 29 1999 |
其他
其他 | The 26th Annual Meeting: Computers in Cardiology 1999 |
---|---|
城市 | Hannover, Ger |
期間 | 9/26/99 → 9/29/99 |
ASJC Scopus subject areas
- 軟體
- 心臟病學與心血管醫學