Abstract
The incidence of breast cancer is increasing worldwide, reinforcing the importance of breast screening. Conventional hand-held ultrasound (HHUS) for breast screening is efficient and relatively easy to perform; however, it lacks systematic recording and localization. This study investigated an electromagnetic tracking-based whole-breast ultrasound (WBUS) system to facilitate the use of HHUS for breast screening. One-hundred nine breast masses were collected, and the detection of suspicious breast lesions was compared between the WBUS system, HHUS and a commercial automated breast ultrasound (ABUS) system. The positioning error between WBUS and ABUS (1.39 ± 0.68 cm) was significantly smaller than that between HHUS and ABUS (1.62 ± 0.91 cm, p = 0.014) and HHUS and WBUS (1.63 ± 0.9 cm, p = 0.024). WBUS is a practical clinical tool for breast screening that can be used instead of the often unavailable and costly ABUS.
Original language | English |
---|---|
Pages (from-to) | 926-933 |
Number of pages | 8 |
Journal | Ultrasound in Medicine and Biology |
Volume | 43 |
Issue number | 5 |
DOIs | |
Publication status | Published - May 1 2017 |
Keywords
- Automated breast ultrasound
- Breast mass
- Comparative study
- Computer-aided tracking system
- Hand-held ultrasound
- Whole-breast ultrasound
ASJC Scopus subject areas
- Radiological and Ultrasound Technology
- Biophysics