TY - JOUR
T1 - The effects of tibia profile, distraction angle, and knee load on wedge instability and hinge fracture
T2 - A finite element study
AU - Weng, Pei Wei
AU - Chen, Chia Hsien
AU - Luo, Chu An
AU - Sun, Jui Sheng
AU - Tsuang, Yang Hwei
AU - Cheng, Cheng Kung
AU - Lin, Shang Chih
N1 - Publisher Copyright:
© 2017 IPEM
PY - 2017/4/1
Y1 - 2017/4/1
N2 - Several plate systems for high tibial osteotomy (HTO) have been developed to stabilize the opening wedge of an osteotomized tibia. Among them, the TomoFix system, having a quasi-straight and T-shaped design, has been widely adopted in the literature. However, this system is implemented by inserting a lag (i.e., cortical) screw through the proximal combi-hole, to deform the plate and pull the distal tibia toward the plate. This process potentially induces plate springback and creates an elastic preload on the osteotomized tibia, especially at the lateral hinge of the distracted wedge. Using the finite-element method, this study aims to investigate the contoured effect of lag-screw application on the biomechanical behavior of the tibia-plate construct. Two tibial profiles (normal and more concave), three distraction angles (6°, 9°, and 12°), and three knee loads (intraoperative: contouring plate; postoperative: weight and nonweight bearing) are systematically varied in this study. The wedge instability and fracture risk at the lateral hinge are chosen as the comparison indices. The results show the necessity of preoperative planning for a precontoured procedure, rather than elastic deformation using a lag screw. Within the intraoperative period, a more concave tibial profile and/or reduced distraction angle (i.e., 6° or 9°) necessitate a higher compressive load to elastically deform the plate, thereby deteriorating the lateral-hinge fracture risk. A precontoured plate is recommended in the case that the proximal tibia is highly concave and the distraction angle is insufficient to stretch the tibial profile.
AB - Several plate systems for high tibial osteotomy (HTO) have been developed to stabilize the opening wedge of an osteotomized tibia. Among them, the TomoFix system, having a quasi-straight and T-shaped design, has been widely adopted in the literature. However, this system is implemented by inserting a lag (i.e., cortical) screw through the proximal combi-hole, to deform the plate and pull the distal tibia toward the plate. This process potentially induces plate springback and creates an elastic preload on the osteotomized tibia, especially at the lateral hinge of the distracted wedge. Using the finite-element method, this study aims to investigate the contoured effect of lag-screw application on the biomechanical behavior of the tibia-plate construct. Two tibial profiles (normal and more concave), three distraction angles (6°, 9°, and 12°), and three knee loads (intraoperative: contouring plate; postoperative: weight and nonweight bearing) are systematically varied in this study. The wedge instability and fracture risk at the lateral hinge are chosen as the comparison indices. The results show the necessity of preoperative planning for a precontoured procedure, rather than elastic deformation using a lag screw. Within the intraoperative period, a more concave tibial profile and/or reduced distraction angle (i.e., 6° or 9°) necessitate a higher compressive load to elastically deform the plate, thereby deteriorating the lateral-hinge fracture risk. A precontoured plate is recommended in the case that the proximal tibia is highly concave and the distraction angle is insufficient to stretch the tibial profile.
KW - Finite element
KW - High tibial osteotomy
KW - Lateral hinge
KW - Precontoured
UR - http://www.scopus.com/inward/record.url?scp=85011588216&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85011588216&partnerID=8YFLogxK
U2 - 10.1016/j.medengphy.2017.01.007
DO - 10.1016/j.medengphy.2017.01.007
M3 - Article
C2 - 28173978
AN - SCOPUS:85011588216
SN - 1350-4533
VL - 42
SP - 48
EP - 54
JO - Medical Engineering and Physics
JF - Medical Engineering and Physics
ER -