Abstract
Attachment of poly(ethylene glycol) to proteins can mask immune epitopes to increase serum half-life, reduce immunogenicity, and enhance in vivo biological efficacy. However, PEGylation mediated epitope-masking may also limit sensitivity and accuracy of traditional ELISA. We previously described an anti-PEG-based sandwich ELISA for universal assay of PEGylated molecules. Here, we compared the quantitative assessment of PEGylated interferons by anti-PEG and traditional anti-interferon sandwich ELISA. The detection limits for PEG-Intron (12k-PEG) and Pegasys (40k-PEG) were 1.9 and 0.03 ng/mL for anti-PEG ELISA compared to 0.18 and 0.42 ng/mL for traditional anti-interferon sandwich ELISA. These results indicate that the anti-PEG sandwich ELISA was insensitive to PEGylation mediated epitope-masking and the sensitivity increased in proportion to the length of PEG. By contrast, PEG-masking interfered with detection by traditional anti-interferon sandwich ELISA. Human and mouse serum did not affect the sensitivity of anti-PEG ELISA but impeded traditional anti-interferon sandwich ELISA. The anti-PEG sandwich ELISA was comparable to anti-interferon sandwich ELISA and radioassay of 131I-Pegasys in pharmacokinetic studies in mice. The anti-PEG sandwich ELISA provides a sensitive, accurate, and convenient quantitative measurement of PEGylated protein drugs.
Original language | English |
---|---|
Pages (from-to) | 1408-1413 |
Number of pages | 6 |
Journal | Bioconjugate Chemistry |
Volume | 24 |
Issue number | 8 |
DOIs | |
Publication status | Published - Aug 21 2013 |
ASJC Scopus subject areas
- Biotechnology
- Bioengineering
- Biomedical Engineering
- Pharmacology
- Pharmaceutical Science
- Organic Chemistry