TY - JOUR
T1 - Mass screening of suspected febrile patients with remote-sensing infrared thermography
T2 - Alarm temperature and optimal distance
AU - Chiang, Ming Fu
AU - Lin, Po Wei
AU - Lin, Li Fong
AU - Chiou, Hung Yi
AU - Chien, Ching Wen
AU - Chu, Shu Fen
AU - Chiu, Wen Ta
N1 - Funding Information:
This study was supported by a grant from the Collaboration Research Fund, Mackay Memorial Hospital, Taipei Medical University, Taipei, Taiwan. We also acknowledge the Chen-Han Foundation for Education.
PY - 2008/12
Y1 - 2008/12
N2 - Background/Purpose: Detection of fever has become an essential step in identifying patients who may have severe acute respiratory syndrome (SARS) or avian influenza. This study evaluated infrared thermography (IRT) and compared the influence of different imagers, ambient temperature discrepancy, and the distance between the subject and imager. Methods: IRT-digital infrared thermal imaging (IRT-DITI), thermoguard, and ear drum IRT were used for visitors to Municipal Wang Fang Hospital, Taipei, Taiwan. The McNemar and χ2 test, standard Pearson correlation, ANOVA, intraclass correlation coefficient (ICC), and receiver operating characteristic curve (ROC) analysis were used to calculate the alarm temperature for each imager. Results: A total of 1032 subjects were recruited. Different distances and ambient temperature discrepancy had a significant influence on thermoguard, and lateral and frontal view DITI By ICC analysis, a significant difference was found at 10 m distance between ear drum IRT and thermoguard (r=0.45), lateral view DITI (r=0.37), and frontal view DITI (r=0.44). With ROC analysis, the optimal preset cut-off temperatures for the different imagers were: 36.05°C for thermoguard (area under the curve [AUC], 0.716), 36.25°C for lateral view DITI (AUC, 0.801), and 36.25°C for frontal view DITI (AUC, 0.812). Conclusion: The temperature readings obtained by IRT may be used as a proxy for core temperature. An effective IRT system with a strict operating protocol can be rapidly implemented at the entrance of a hospital during SARS or avian influenza epidemics.
AB - Background/Purpose: Detection of fever has become an essential step in identifying patients who may have severe acute respiratory syndrome (SARS) or avian influenza. This study evaluated infrared thermography (IRT) and compared the influence of different imagers, ambient temperature discrepancy, and the distance between the subject and imager. Methods: IRT-digital infrared thermal imaging (IRT-DITI), thermoguard, and ear drum IRT were used for visitors to Municipal Wang Fang Hospital, Taipei, Taiwan. The McNemar and χ2 test, standard Pearson correlation, ANOVA, intraclass correlation coefficient (ICC), and receiver operating characteristic curve (ROC) analysis were used to calculate the alarm temperature for each imager. Results: A total of 1032 subjects were recruited. Different distances and ambient temperature discrepancy had a significant influence on thermoguard, and lateral and frontal view DITI By ICC analysis, a significant difference was found at 10 m distance between ear drum IRT and thermoguard (r=0.45), lateral view DITI (r=0.37), and frontal view DITI (r=0.44). With ROC analysis, the optimal preset cut-off temperatures for the different imagers were: 36.05°C for thermoguard (area under the curve [AUC], 0.716), 36.25°C for lateral view DITI (AUC, 0.801), and 36.25°C for frontal view DITI (AUC, 0.812). Conclusion: The temperature readings obtained by IRT may be used as a proxy for core temperature. An effective IRT system with a strict operating protocol can be rapidly implemented at the entrance of a hospital during SARS or avian influenza epidemics.
KW - Avian influenza
KW - Infrared thermography
KW - Intraclass correlation coefficient
KW - Receiver operating characteristic curve
KW - Severe acute respiratory syndrome
UR - http://www.scopus.com/inward/record.url?scp=59149089907&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=59149089907&partnerID=8YFLogxK
U2 - 10.1016/S0929-6646(09)60017-6
DO - 10.1016/S0929-6646(09)60017-6
M3 - Article
C2 - 19129054
AN - SCOPUS:59149089907
SN - 0929-6646
VL - 107
SP - 937
EP - 944
JO - Journal of the Formosan Medical Association
JF - Journal of the Formosan Medical Association
IS - 12
ER -