Skip to main navigation Skip to search Skip to main content

Fracture of a perfusion balloon catheter during coronary angioplasty

Research output: Contribution to journalArticlepeer-review

Abstract

In a 74‐year‐old patient with increasing angina, an attempt was made to dilate an ostial stenosis in the first diagonal artery. The initial attempt with a long balloon catheter was unsuccessful. Because of chest pain associated with hypotension during balloon inflations, a second attempt with a perfusion balloon catheter resulted in fracture of this catheter's tip (14 mm long) in the artery. Bypass graft surgery was performed and the retained fragment was easily removed by arteriotomy. The patient made an uneventful recovery. © 1994 Wiley‐Liss,Inc..

Original languageEnglish
Pages (from-to)33-35
Number of pages3
JournalCatheterization and Cardiovascular Diagnosis
Volume32
Issue number1
DOIs
Publication statusPublished - Jan 1 1994
Externally publishedYes

UN SDGs

This output contributes to the following UN Sustainable Development Goals (SDGs)

  1. SDG 3 - Good Health and Well-being
    SDG 3 Good Health and Well-being

Keywords

  • arteriotomy
  • perfusion balloon catheter
  • PTCA complication

ASJC Scopus subject areas

  • Cardiology and Cardiovascular Medicine

Fingerprint

Dive into the research topics of 'Fracture of a perfusion balloon catheter during coronary angioplasty'. Together they form a unique fingerprint.

Cite this