TY - JOUR
T1 - Factors in ProGlide® Vascular Closure Failure in Sheath Arteriotomies Greater than 16 French
AU - Chen, I-Ming
AU - Lee, Tsung-Hsing
AU - Chen, Po-Lin
AU - Shih, Chun-Che
AU - Chang, Hsiao-Hunag
N1 - Copyright © 2019 European Society for Vascular Surgery. Published by Elsevier B.V. All rights reserved.
PY - 2019/10
Y1 - 2019/10
N2 - OBJECTIVES: The ProGlide® vascular closure device (Abbott Vascular, Redwood City, CA, USA) is approved for the closure of arterial punctures (typically 5-21 Fr sheath; maximum outer diameter, 26 Fr). However, a failure rate of about 2-8% is reported. This study was conducted to analyse factors predisposing to failure when the devices were used for the closure of large hole (16-26 Fr) arteriotomies, and to determine the predictive cut off values of predisposing factors.METHODS: In this retrospective study, the ProGlide® device was used to achieve vascular access site closure in 458 patients undergoing repair of abdominal aortic aneurysm, thoracic aortic aneurysm, type B aortic dissection, or transcatheter aortic valve implantation. The primary endpoint was device failure, defined as inability to achieve common femoral artery (CFA) closure; successful repair, development of acute lower limb ischaemia and haemodynamic compromise; or delayed pseudoaneurysm formation during the follow up period, requiring open repair.RESULTS: Overall, ProGlide® failure occurred in 7.6% of cases. Factors that predisposed to failure included a history of peripheral arterial disease (PAD) (p < .001), the presence of CFA calcification (p < .001), the depth of the skin puncture site ≥ 33 mm (p < .001), body mass index (BMI) of ≥28.7 kg/m2 (p < .001), and use of sheath size ≥ 19 Fr (p < .001).CONCLUSION: Factors such as BMI, history of PAD, the presence of CFA calcification, the depth of the skin puncture site, and sheath size are significantly associated with ProGlide® failure. Hence, careful patient and device selection and operating procedure are paramount to achieve successful outcomes.
AB - OBJECTIVES: The ProGlide® vascular closure device (Abbott Vascular, Redwood City, CA, USA) is approved for the closure of arterial punctures (typically 5-21 Fr sheath; maximum outer diameter, 26 Fr). However, a failure rate of about 2-8% is reported. This study was conducted to analyse factors predisposing to failure when the devices were used for the closure of large hole (16-26 Fr) arteriotomies, and to determine the predictive cut off values of predisposing factors.METHODS: In this retrospective study, the ProGlide® device was used to achieve vascular access site closure in 458 patients undergoing repair of abdominal aortic aneurysm, thoracic aortic aneurysm, type B aortic dissection, or transcatheter aortic valve implantation. The primary endpoint was device failure, defined as inability to achieve common femoral artery (CFA) closure; successful repair, development of acute lower limb ischaemia and haemodynamic compromise; or delayed pseudoaneurysm formation during the follow up period, requiring open repair.RESULTS: Overall, ProGlide® failure occurred in 7.6% of cases. Factors that predisposed to failure included a history of peripheral arterial disease (PAD) (p < .001), the presence of CFA calcification (p < .001), the depth of the skin puncture site ≥ 33 mm (p < .001), body mass index (BMI) of ≥28.7 kg/m2 (p < .001), and use of sheath size ≥ 19 Fr (p < .001).CONCLUSION: Factors such as BMI, history of PAD, the presence of CFA calcification, the depth of the skin puncture site, and sheath size are significantly associated with ProGlide® failure. Hence, careful patient and device selection and operating procedure are paramount to achieve successful outcomes.
U2 - 10.1016/j.ejvs.2019.03.037
DO - 10.1016/j.ejvs.2019.03.037
M3 - Article
C2 - 31500989
SN - 1078-5884
VL - 58
SP - 615
EP - 622
JO - European Journal of Vascular and Endovascular Surgery
JF - European Journal of Vascular and Endovascular Surgery
IS - 4