TY - JOUR
T1 - Evaluation of loading conditions on fatigue-failed implants by fracture surface analysis
AU - Huang, H.-M.
AU - Tsai, C.-M.
AU - Chang, C.-C.
AU - Lin, C.-T.
AU - Lee, S.-Y.
N1 - 被引用次數:15
Export Date: 9 August 2016
通訊地址: Lee, S.-Y.; School of Dentistry, Taipei Medical University, 250 Wu-Hsing Street, Taipei, Taiwan; 電子郵件: [email protected]
化學物質/CAS: Dental Implants
參考文獻: Jemt, T., Failure and complications in 391 consecutive inserted fixed prostheses supported by Brånemark implants in edentulous jaws: A study of treatment from the time of prosthesis placement to the first annual checkup (1991) Int J Oral Maxillofac Implants, 6, pp. 270-276; Lekholm, U., Van Steenberghe, D., Herrmann, I., Osseointegrated implants in the treatment of partially edentulous jaws: A prospective 5-year multicenter study (1994) Int J Oral Maxillofac Implants, 9, pp. 627-635; Jemt, T., Linden, B., Lekholm, U., Failures and complications in 127 consecutively placed fixed partial prostheses supported by Brånemark implants: From prosthetic treatment to first annual checkup (1992) Int J Oral Maxillofac Implants, 7, pp. 40-44; Hoyer, H.A., Stanford, C.M., Buranadham, S., Fridrich, T., Wagner, J., Gratton, D., Dynamic fatigue properties of the dental implant-abutment interface: Joint opening in wide-diameter versus standard-diameter hex-type implants (2001) J Prosthet Dent, 85, pp. 599-607; Versluis, A., Korioth, T.W., Cardoso, A.C., Numerical analysis of a dental implant system preloaded with a washer (1999) Int J Oral Maxillofac Implants, 14, pp. 337-341; Sakaguchi, R.L., Borgersen, S.E., Nonlinear contact analysis of preload in dental implant screws (1995) Int J Oral Maxillofac Implants, 10, pp. 295-302; Balfour, A., O'Brien, G.R., Comparative study of antirotational single tooth abutments (1995) J Prosthet Dent, 73, pp. 36-43; Basten, C.H., Nicholls, J.I., Daly, C.H., Taggart, R., Load fatigue performance of two implant-abutment combinations (1996) Int J Oral Maxillofac Implants, 11, pp. 522-528; Boggan, R.S., Strong, J.T., Misch, C.E., Bidez, M.W., Influence of hex geometry and prosthetic table width on static and fatigue strength of dental implants (1999) J Prosthet Dent, 82, pp. 436-440; Kelly, J.R., Campbell, S.D., Bowen, H.K., Fracture-surface analysis of dental ceramics (1989) J Prosthet Dent, 62, pp. 536-541; Thompson, J.Y., Anusavice, K.J., Naman, A., Morris, H.F., Fracture surface characterization of clinically failed all-ceramic crowns (1994) J Dent Res, 73, pp. 1824-1832; Mecholsky Jr., J.J., Fracture mechanics principles (1995) Dent Mater, 11, pp. 111-112; Mecholsky Jr., J.J., Fractography: Determining the sites of fracture initiation (1995) Dent Mater, 11, pp. 113-116; Vallittu, P.K., Kokkonen, M., Deflection fatigue of cobalt-chromium, titanium, and gold alloy cast denture clasp (1995) J Prosthet Dent, 74, pp. 412-419; Vallittu, P.K., Fracture surface characteristics of damaged acrylic-resin-based dentures as analyzed by SEM - Replica technique (1996) J Oral Rehabil, 23, pp. 524-529; Vallittu, P.K., Brazing joints of gold alloy used in porcelain-fused-to-metal restorations and their resistance to deflection fatigue (1997) J Oral Rehabil, 24, pp. 444-448; Morgan, M.J., James, D.F., Pilliar, R.M., Fractures of the fixture component of an osseointegrated implant (1993) Int J Oral Maxillofac Implants, 8, pp. 409-414; Binon, P.P., McHugh, M.J., The effect of eliminating implant/abutment relational misfit on screw joint stability (1996) Int J Prosthodont, 9, pp. 511-519; Bidez, M.W., Misch, C.E., Clinical biomechanics in implant dentistry (1999) Contemporary Implant Dentistry, Ed 2, pp. 303-316. , Misch CE (ed). St Louis: CV Mosby; Gibbs, C.H., Mahan, P.E., Lundeen, H.C., Brehran, K., Walsh, E.K., Holbrook, W.B., Occlusal forces during chewing and swallowing as measured by sound transmission (1981) J Prosthet Dent, 46, pp. 443-449; Wiskott, H.W.A., Nicholls, J.I., Belser, U.C., Stress fatigue: Basic principles and prosthodontic implications (1995) Int J Prosthodont, 8, pp. 105-116; Soncini, M., Pietrabissa, R.P., Natali, A.N., Pavan, P.G., Williams, K.R., Testing the reliability of dental implant devices (2003) Dental Biomechanics, pp. 120-122. , Natali AN (ed). New York: Taylor & Francis; Van Vlack, L.H., Delayed fracture (1989) Elements of Materials Science and Engineering, pp. 520-527. , Van Vlack LH. (ed). Reading, MA: Addison-Wesley
PY - 2005
Y1 - 2005
N2 - Purpose: The goal of this study was to determine the relationship between fracture surface morphology and applied stress level for dental abutment screws loaded in cyclic fatigue. If a correlation between fracture surface and load level can be determined, then the fracture surface analysis could be used as a tool to assess the mechanism by which a screw failed and the magnitude of the load at which it failed. Materials and Methods: Test implants were loaded with static and cyclic forces. In the cyclic test, the load versus the number of cycles was plotted as a curve for biomechanical analysis. The fracture surfaces of the failed screws were observed and recorded using scanning electron microscopy (SEM). Results: Two fracture phases, a smooth region and a rough region, were observed on the fracture surface. After identifying the boundary between the 2 regions, the smooth region ratio (SRR), the ratio of the smooth phase area to the area of the whole fracture surface, was measured using digitized SEM images. The mean SRRs were 0.60 ± 0.03, 0.66 ± 0.03, and 0.75 ± 0.03 when the tested implants were subjected to dynamic loading of 60%, 55%, and 50% ultimate failure loading (UFL), respectively. Linear relationships were found between the SRR values and loading magnitude and between SSR and number of cycles. Discussion: The smooth area on the fracture surface can be used to assess the load conditions and internal stress of fatigue-fractured implants. Conclusions: These results demonstrate that fracture surface analysis of fractured implants has the potential to become a useful indicator for assessing implant fracture mechanisms.
AB - Purpose: The goal of this study was to determine the relationship between fracture surface morphology and applied stress level for dental abutment screws loaded in cyclic fatigue. If a correlation between fracture surface and load level can be determined, then the fracture surface analysis could be used as a tool to assess the mechanism by which a screw failed and the magnitude of the load at which it failed. Materials and Methods: Test implants were loaded with static and cyclic forces. In the cyclic test, the load versus the number of cycles was plotted as a curve for biomechanical analysis. The fracture surfaces of the failed screws were observed and recorded using scanning electron microscopy (SEM). Results: Two fracture phases, a smooth region and a rough region, were observed on the fracture surface. After identifying the boundary between the 2 regions, the smooth region ratio (SRR), the ratio of the smooth phase area to the area of the whole fracture surface, was measured using digitized SEM images. The mean SRRs were 0.60 ± 0.03, 0.66 ± 0.03, and 0.75 ± 0.03 when the tested implants were subjected to dynamic loading of 60%, 55%, and 50% ultimate failure loading (UFL), respectively. Linear relationships were found between the SRR values and loading magnitude and between SSR and number of cycles. Discussion: The smooth area on the fracture surface can be used to assess the load conditions and internal stress of fatigue-fractured implants. Conclusions: These results demonstrate that fracture surface analysis of fractured implants has the potential to become a useful indicator for assessing implant fracture mechanisms.
KW - Dental implants
KW - Fatigue
KW - Finite element analysis
KW - Implant fractures
KW - Scanning electron micrography
KW - article
KW - dental surgery
KW - equipment
KW - finite element analysis
KW - mechanical stress
KW - scanning electron microscopy
KW - surface property
KW - tooth implantation
KW - weight bearing
KW - Dental Implants
KW - Dental Restoration Failure
KW - Equipment Failure Analysis
KW - Finite Element Analysis
KW - Microscopy, Electron, Scanning
KW - Stress, Mechanical
KW - Surface Properties
KW - Weight-Bearing
UR - https://www.scopus.com/record/display.uri?eid=2-s2.0-30844453838&origin=inward&txGid=28f45b0961c118a7f2371b3d22004e69
UR - https://www.scopus.com/results/citedbyresults.uri?sort=plf-f&cite=2-s2.0-30844453838&src=s&imp=t&sid=de5d5ac3eb5571cc80e301fee2baf40a&sot=cite&sdt=a&sl=0&origin=recordpage&editSaveSearch=&txGid=35a3d98ab053b382e043d0bd0fee363a
M3 - Article
SN - 0882-2786
VL - 20
SP - 854
EP - 859
JO - International Journal of Oral and Maxillofacial Implants
JF - International Journal of Oral and Maxillofacial Implants
IS - 6
ER -