Effects of the electrode location on tonal discrimination and speech perception of mandarin-speaking patients with a cochlear implant

Fei Peng Lee, Hsin Te Hsu, Yong-Song Lin, Su Chen Hung

Research output: Contribution to journalArticlepeer-review

5 Citations (Scopus)

Abstract

Objectives/Hypothesis: This study assessed the effects of varying the electrode location on tonal discrimination and speech perception in Mandarin Chinese-speaking patients. Study Design: A controlled study with six experimental conditions. Methods: Seven Mandarin-speaking listeners who received a MED-EL cochlear implant (CI), ranging in age from 12.88 to 36.43 years (mean, 25.51 years), with an average of 5.28 years of device experience, participated this study. To evaluate the effects of electrode location, six experimental conditions each with the switch off at six different electrodes were designed. Identification tests of Mandarin lexical tones and words were performed. Results: Among experimental conditions with electrode lengths of 31, 23.8, and 16.6 mm, the CI subjects exhibited improved vowel and consonant identification in the condition of 31 mm, reflecting the apical location of electrodes. Specifically, the improvement was observed in the identification score for the vowel backness and height, as well as for the consonant place of articulation. Comparison among three settings with a same electrode length of 12.6 mm and the setting with stimulation to the midregion of the cochlea produces better words as well as the vowel and consonant identification compared with stimulation to basal and apical regions. However, no significant difference was observed for the lexical tone identification among conditions with different electrode location and stimulating region. Conclusions: Less mismatch of the frequency-to-place alignment may account for the improvement of word identification in conditions with electrodes coverage to more apical location; and in conditions where the mid-region of the cochlea were stimulated. Laryngoscope, 2012

Original languageEnglish
Pages (from-to)1366-1378
Number of pages13
JournalLaryngoscope
Volume122
Issue number6
DOIs
Publication statusPublished - Jun 2012

Keywords

  • Cochlear implant
  • auditory physiology
  • hearing loss
  • lexical tone

ASJC Scopus subject areas

  • Otorhinolaryngology

Fingerprint

Dive into the research topics of 'Effects of the electrode location on tonal discrimination and speech perception of mandarin-speaking patients with a cochlear implant'. Together they form a unique fingerprint.

Cite this