T1 - Diagnostic accuracy of ambulatory devices in detecting atrial fibrillation
T2 - Systematic review and meta-analysis
AU - Yang, Tien Yun
AU - Huang, Li
AU - Malwade, Shwetambara
AU - Hsu, Chien Yi
AU - Chen, Yang Ching
N1 - Funding Information:
The authors thank Mark Lown, Felix K Wegner, and Aviram Hochstadt for sharing their original data and contributing to the analytic calculations. The authors also thank Wallace Academic Editing for manuscript editing. We also thank the Higher Education Sprout Project of the Ministry of Education (MOE) in Taiwan for financially supporting this work [DP2-110-21121-01-A-04].
Publisher Copyright:
© 2021 Tien Yun Yang, Li Huang, Shwetambara Malwade, Chien-Yi Hsu, Yang Ching Chen.
PY - 2021/4
Y1 - 2021/4
N2 - Background: Atrial fibrillation (AF) is the most common cardiac arrhythmia worldwide. Early diagnosis of AF is crucial for preventing AF-related morbidity, mortality, and economic burden, yet the detection of the disease remains challenging. The 12-lead electrocardiogram (ECG) is the gold standard for the diagnosis of AF. Because of technological advances, ambulatory devices may serve as convenient screening tools for AF. Objective: The objective of this review was to investigate the diagnostic accuracy of 2 relatively new technologies used in ambulatory devices, non-12-lead ECG and photoplethysmography (PPG), in detecting AF. We performed a meta-analysis to evaluate the diagnostic accuracy of non-12-lead ECG and PPG compared to the reference standard, 12-lead ECG. We also conducted a subgroup analysis to assess the impact of study design and participant recruitment on diagnostic accuracy. Methods: This systematic review and meta-analysis was conducted in accordance with the Preferred Reporting Items for Systematic Reviews and Meta-Analysis (PRISMA) guidelines. MEDLINE and EMBASE were systematically searched for articles published from January 1, 2015 to January 23, 2021. A bivariate model was used to pool estimates of sensitivity, specificity, positive likelihood ratio (PLR), negative likelihood ratio (NLR), and area under the summary receiver operating curve (SROC) as the main diagnostic measures. Study quality was evaluated using the quality assessment of diagnostic accuracy studies (QUADAS-2) tool. Results: Our search resulted in 16 studies using either non-12-lead ECG or PPG for detecting AF, comprising 3217 participants and 7623 assessments. The pooled estimates of sensitivity, specificity, PLR, NLR, and diagnostic odds ratio for the detection of AF were 89.7% (95% CI 83.2%-93.9%), 95.7% (95% CI 92.0%-97.7%), 20.64 (95% CI 10.10-42.15), 0.11 (95% CI 0.06-0.19), and 224.75 (95% CI 70.10-720.56), respectively, for the automatic interpretation of non-12-lead ECG measurements and 94.7% (95% CI 93.3%-95.8%), 97.6% (95% CI 94.5%-99.0%), 35.51 (95% CI 18.19-69.31), 0.05 (95% CI 0.04-0.07), and 730.79 (95% CI 309.33-1726.49), respectively, for the automatic interpretation of PPG measurements. Conclusions: Both non-12-lead ECG and PPG offered high diagnostic accuracies for AF. Detection employing automatic analysis techniques may serve as a useful preliminary screening tool before administering a gold standard test, which generally requires competent physician analyses. Subgroup analysis indicated variations of sensitivity and specificity between studies that recruited low-risk and high-risk populations, warranting future validity tests in the general population.
AB - Background: Atrial fibrillation (AF) is the most common cardiac arrhythmia worldwide. Early diagnosis of AF is crucial for preventing AF-related morbidity, mortality, and economic burden, yet the detection of the disease remains challenging. The 12-lead electrocardiogram (ECG) is the gold standard for the diagnosis of AF. Because of technological advances, ambulatory devices may serve as convenient screening tools for AF. Objective: The objective of this review was to investigate the diagnostic accuracy of 2 relatively new technologies used in ambulatory devices, non-12-lead ECG and photoplethysmography (PPG), in detecting AF. We performed a meta-analysis to evaluate the diagnostic accuracy of non-12-lead ECG and PPG compared to the reference standard, 12-lead ECG. We also conducted a subgroup analysis to assess the impact of study design and participant recruitment on diagnostic accuracy. Methods: This systematic review and meta-analysis was conducted in accordance with the Preferred Reporting Items for Systematic Reviews and Meta-Analysis (PRISMA) guidelines. MEDLINE and EMBASE were systematically searched for articles published from January 1, 2015 to January 23, 2021. A bivariate model was used to pool estimates of sensitivity, specificity, positive likelihood ratio (PLR), negative likelihood ratio (NLR), and area under the summary receiver operating curve (SROC) as the main diagnostic measures. Study quality was evaluated using the quality assessment of diagnostic accuracy studies (QUADAS-2) tool. Results: Our search resulted in 16 studies using either non-12-lead ECG or PPG for detecting AF, comprising 3217 participants and 7623 assessments. The pooled estimates of sensitivity, specificity, PLR, NLR, and diagnostic odds ratio for the detection of AF were 89.7% (95% CI 83.2%-93.9%), 95.7% (95% CI 92.0%-97.7%), 20.64 (95% CI 10.10-42.15), 0.11 (95% CI 0.06-0.19), and 224.75 (95% CI 70.10-720.56), respectively, for the automatic interpretation of non-12-lead ECG measurements and 94.7% (95% CI 93.3%-95.8%), 97.6% (95% CI 94.5%-99.0%), 35.51 (95% CI 18.19-69.31), 0.05 (95% CI 0.04-0.07), and 730.79 (95% CI 309.33-1726.49), respectively, for the automatic interpretation of PPG measurements. Conclusions: Both non-12-lead ECG and PPG offered high diagnostic accuracies for AF. Detection employing automatic analysis techniques may serve as a useful preliminary screening tool before administering a gold standard test, which generally requires competent physician analyses. Subgroup analysis indicated variations of sensitivity and specificity between studies that recruited low-risk and high-risk populations, warranting future validity tests in the general population.
KW - Ambulatory device
KW - Ambulatory devices
KW - Atrial fibrillation
KW - Diagnostic accuracy
KW - Electrocardiogram
KW - Mobile health
KW - Photoplethysmography
KW - Technology
KW - Ubiquitous health
UR - http://www.scopus.com/inward/record.url?scp=85104148711&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85104148711&partnerID=8YFLogxK
U2 - 10.2196/26167
DO - 10.2196/26167
M3 - Review article
C2 - 33835039
AN - SCOPUS:85104148711
SN - 2291-5222
VL - 9
JO - JMIR mHealth and uHealth
JF - JMIR mHealth and uHealth
IS - 4
M1 - e26167
ER -