Corrigendum: Advances of techniques in deep regional blocks (BioMed Research International (2017) 2017 (4) DOI: 10.1155/2017/7268308)

Jui An Lin, Rafael Blanco, Yasuyuki Shibata, Tatsuo Nakamoto, Ko Huan Lin

Research output: Contribution to journalComment/debatepeer-review

1 Citation (Scopus)

Abstract

In the article titled "Advances of Techniques in Deep Regional Blocks" [1], Table 1 was missing. Accordingly, the table is shown below, and its in-text citation is added as follows: Our preliminary qualitative analysis demonstrated that, by using the pressure management system of Injectomat Agilia® pump (Fresenius Vial, Brezins, France) as an inline manometer between the needle (the tip inserted 3 cm into the pork model) and the low-dead space extension tube, pushing pressure generated by the act of half-the-air was below 15 psi during injection (experiment was run in triplicate, and occlusion alarm did not occur after the flow had commenced in response to half-the-air pressure exerted in the 20mL local anesthetic syringe with the syringe pump set to an infusion rate of 0.1ml/h and a pressure limit of 750mmHg) (Table 1). Table presented.

Original languageEnglish
Article number5151645
JournalBioMed Research International
Volume2018
DOIs
Publication statusPublished - Jan 1 2018